会议论文详细信息
| 8th Symposium on Frequency Standards and Metrology 2015 | |
| Microfabricated Optically-Pumped Magnetometers for Biomagnetic Applications | |
| Knappe, Svenja^1 ; Alem, Orang^1 ; Sheng, Dong^1 ; Kitching, John^1 | |
| NIST, 325 Broadway, Boulder | |
| CO | |
| 80305, United States^1 | |
| 关键词: Biomedical measurements; Heart activities; Imaging applications; Magnetic field noise; Magnetometer arrays; Microfabricated; Optically pumped; Orders of magnitude; | |
| Others : https://iopscience.iop.org/article/10.1088/1742-6596/723/1/012055/pdf DOI : 10.1088/1742-6596/723/1/012055 |
|
| 来源: IOP | |
PDF
|
|
【 摘 要 】
We report on the progress in developing microfabricated optically-pumped magnetometer arrays for imaging applications. We have improved our sensitivities by several orders of magnitude in the last ten years. Now, our zero-field magnetometers reach noise values below 15 fT/Hz1/2. Recently, we have also developed gradiometers to reject ambient magnetic field noise. We have built several imaging arrays and validated them for biomedical measurements of brain and heart activity.
【 预 览 】
| Files | Size | Format | View |
|---|---|---|---|
| Microfabricated Optically-Pumped Magnetometers for Biomagnetic Applications | 1106KB |
PDF