8th Symposium on Frequency Standards and Metrology 2015 | |
Measurement of the Microwave Lensing shift in NIST-F1 and NIST-F2 | |
Jefferts, S.R.^1 ; Heavner, T.P.^1 ; Barlow, S.E.^1 ; Ashby, N.^1 | |
NIST Time and Frequency Division, National Institute for Standards and Technology, Boulder | |
CO | |
80305, United States^1 | |
关键词: Coordinated universal time; Fractional frequencies; Frequency shift; Measurements of; Primary frequency standard; | |
Others : https://iopscience.iop.org/article/10.1088/1742-6596/723/1/012005/pdf DOI : 10.1088/1742-6596/723/1/012005 |
|
来源: IOP | |
【 摘 要 】
With several Primary Frequency Standards (PFS) across the world demonstrating systematic fractional frequency uncertainties on order of 1 x 10-16, it is crucial to accurately measure or model even small frequency shifts that could affect the ultimate PFS uncertainty, and thus ultimately impact the rate of Coordinated Universal Time (UTC) which relies on precision PFS measurements. Recently there has been controversy about the physical causes and size of PFS frequency shifts due to microwave lensing effects. We present here the first measurements of microwave lensing frequency shifts in the PFS NIST-F1 and NIST-F2. The measured frequency shifts agree well with the recent theory of Ashby et al [1].
【 预 览 】
Files | Size | Format | View |
---|---|---|---|
Measurement of the Microwave Lensing shift in NIST-F1 and NIST-F2 | 622KB | download |