14th International Conference on Topics in Astroparticle and Underground Physics | |
Using 220Rn to calibrate liquid noble gas detectors | |
物理学;天文学 | |
Kobayashi, M.^1 ; Yamashita, M.^1 ; Takeda, A.^1 ; Kishimoto, K.^1 ; Moriyama, S.^1 | |
Kamioka Observatory, Institute for Cosmic Ray Research, University of Tokyo, Higashi-Mozumi, Kamioka, Hida, Gifu | |
506-1205, Japan^1 | |
关键词: Calibration source; Carrier gas; Hot metal; Liquid noble gas; Liquid xenon; Low-energy; Test chambers; Test Environment; | |
Others : https://iopscience.iop.org/article/10.1088/1742-6596/718/4/042069/pdf DOI : 10.1088/1742-6596/718/4/042069 |
|
学科分类:天文学(综合) | |
来源: IOP | |
【 摘 要 】
In this paper, we describe220Rn calibration source that was developed for liquid noble gas detectors. The key advantage of this source is that it can provide212Bi-212Po consecutive events, which enables us to evaluate the vertex resolution of a detector at low energy by comparing low-energy events of212Bi and corresponding higher-energy α-rays from212Po. Since220Rn is a noble gas, a hot metal getter can be used when introduced using xenon as the carrier gas. In addition, no long-life radioactive isotopes are left behind in the detector after the calibration is complete; this has clear advantage over the use of222Rn which leaves longlife radioactivity, i.e.,210Pb. Using a small liquid xenon test chamber, we developed a system to introduce220Rn via the xenon carrier gas; we demonstrated the successful introduction of 6 ×102220Rn atoms in our test environment.