Geant4 simulation for a study of a possible use of carbon ion pencil beams for the treatment of ocular melanomas with the active scanning system at CNAO
物理学;计算机科学
Farina, E.^1 ; Piersimoni, P.^2 ; Riccardi, C.^1,3 ; Rimoldi, A.^1,3 ; Tamborini, A.^1,3 ; Ciocca, M.^4
University of Pavia, Department of Physics, via Bassi 6, Pavia
27100, Italy^1
Division of Radiation Research, Loma Linda University, Loma Linda
CA
92354, United States^2
INFN Section of Pavia, via Bassi 6, Pavia
27100, Italy^3
Medical Physics Unit, Centro Nazionale di Adroterapia Oncologica (CNAO Foundation), Strada Campeggi 53, Pavia
27100, Italy^4
关键词: Active scanning;    Biological effectiveness;    Carbon ion beams;    Carbon ions;    GEANT4 simulation;    Pencil beam;    Transverse distribution;    Tumor volumes;   
Others  :  https://iopscience.iop.org/article/10.1088/1742-6596/664/7/072048/pdf
DOI  :  10.1088/1742-6596/664/7/072048
学科分类:计算机科学(综合)
来源: IOP
PDF
【 摘 要 】
The aim of this work was to study a possible use of carbon ion pencil beams (delivered with active scanning modality) for the treatment of ocular melanomas at the Centro Nazionale di Adroterapia Oncologica (CNAO). The promising aspect of carbon ions radiotherapy for the treatment of this disease lies in its superior relative radio-biological effectiveness (RBE). The Monte Carlo (MC) Geant4 10.00 toolkit was used to simulate the complete CNAO extraction beamline, with the active and passive components along it. A human eye modeled detector, including a realistic target tumor volume, was used as target. Cross check with previous studies at CNAO using protons allowed comparisons on possible benefits on using such a technique with respect to proton beams. Experimental data on proton and carbon ion beams transverse distributions were used to validate the simulation.