会议论文详细信息
International Conference on Bio-Medical Instrumentation and related Engineering and Physical Sciences | |
Image Quality Assessment of a CMOS/Gd2O2S:Pr,Ce,F X-ray Sensor | |
物理学;医药卫生 | |
Michail, C.M.^1 ; Seferis, I.E.^2 ; Sideras, T.^1 ; Valais, I.G.^1 ; Fountos, G.P.^1 ; Bakas, A.^3 ; Panayiotakis, G.S.^4 ; Kandarakis, I.S.^1 | |
Radiation Physics, Materials Technology and Biomedical Imaging Laboratory, Department of Biomedical Engineering, Technological Educational Institute of Athens, Egaleo, Athens | |
122 10, Greece^1 | |
Faculty of Chemistry, Wroclaw University, 14F Joliot-Curie Street, Wroclaw | |
50-383, Poland^2 | |
Medical Radiological Technology, Faculty of Health and Caring Professions, Technological Educational Institute of Athens, Athens | |
122 10, Greece^3 | |
Department of Medical Physics, Faculty of Medicine, University of Patras, Patras | |
265 00, Greece^4 | |
关键词: Detector response functions; Gadolinium oxysulfides; Image quality assessment; Imaging properties; Inorganic scintillator; Phosphor technologies; Powder scintillator; Quality performance; | |
Others : https://iopscience.iop.org/article/10.1088/1742-6596/637/1/012018/pdf DOI : 10.1088/1742-6596/637/1/012018 |
|
学科分类:卫生学 | |
来源: IOP | |
【 摘 要 】
The aim of the present study was to examine the image quality performance of a CMOS digital imaging optical sensor coupled to custom made gadolinium oxysulfide powder scintillators, doped with praseodymium, cerium and fluorine (Gd2O2S:Pr,Ce,F) screens. The screens, with coating thicknesses 35.7 and 71.2 mg/cm2, were prepared in our laboratory from Gd2O2S:Pr,Ce,F powder (Phosphor Technology, Ltd) by sedimentation on silica substrates and were placed in direct contact with the optical sensor. Image quality was determined through a single index image quality parameter (information capacity). The CMOS sensor/Gd2O2S:Pr,Ce,F screens combinations were irradiated under the RQA-5 (IEC 62220-1) beam quality. The detector response function was linear for the exposure range under investigation. Under the general radiography conditions, both Gd2O2S:Pr,Ce,F screen/CMOS combinations exhibited comparable overall imaging properties, in terms of the information capacity, to previously published scintillators, such as Gd2O2S:Eu.【 预 览 】
Files | Size | Format | View |
---|---|---|---|
Image Quality Assessment of a CMOS/Gd2O2S:Pr,Ce,F X-ray Sensor | 585KB | download |