会议论文详细信息
International Conference on Bio-Medical Instrumentation and related Engineering and Physical Sciences
Image Quality Assessment of a CMOS/Gd2O2S:Pr,Ce,F X-ray Sensor
物理学;医药卫生
Michail, C.M.^1 ; Seferis, I.E.^2 ; Sideras, T.^1 ; Valais, I.G.^1 ; Fountos, G.P.^1 ; Bakas, A.^3 ; Panayiotakis, G.S.^4 ; Kandarakis, I.S.^1
Radiation Physics, Materials Technology and Biomedical Imaging Laboratory, Department of Biomedical Engineering, Technological Educational Institute of Athens, Egaleo, Athens
122 10, Greece^1
Faculty of Chemistry, Wroclaw University, 14F Joliot-Curie Street, Wroclaw
50-383, Poland^2
Medical Radiological Technology, Faculty of Health and Caring Professions, Technological Educational Institute of Athens, Athens
122 10, Greece^3
Department of Medical Physics, Faculty of Medicine, University of Patras, Patras
265 00, Greece^4
关键词: Detector response functions;    Gadolinium oxysulfides;    Image quality assessment;    Imaging properties;    Inorganic scintillator;    Phosphor technologies;    Powder scintillator;    Quality performance;   
Others  :  https://iopscience.iop.org/article/10.1088/1742-6596/637/1/012018/pdf
DOI  :  10.1088/1742-6596/637/1/012018
学科分类:卫生学
来源: IOP
PDF
【 摘 要 】
The aim of the present study was to examine the image quality performance of a CMOS digital imaging optical sensor coupled to custom made gadolinium oxysulfide powder scintillators, doped with praseodymium, cerium and fluorine (Gd2O2S:Pr,Ce,F) screens. The screens, with coating thicknesses 35.7 and 71.2 mg/cm2, were prepared in our laboratory from Gd2O2S:Pr,Ce,F powder (Phosphor Technology, Ltd) by sedimentation on silica substrates and were placed in direct contact with the optical sensor. Image quality was determined through a single index image quality parameter (information capacity). The CMOS sensor/Gd2O2S:Pr,Ce,F screens combinations were irradiated under the RQA-5 (IEC 62220-1) beam quality. The detector response function was linear for the exposure range under investigation. Under the general radiography conditions, both Gd2O2S:Pr,Ce,F screen/CMOS combinations exhibited comparable overall imaging properties, in terms of the information capacity, to previously published scintillators, such as Gd2O2S:Eu.
【 预 览 】
附件列表
Files Size Format View
Image Quality Assessment of a CMOS/Gd2O2S:Pr,Ce,F X-ray Sensor 585KB PDF download
  文献评价指标  
  下载次数:22次 浏览次数:39次