Response function simulation of the anti-coincidence detector based on NaI crystal with a complex shape in registration systems for the experiments SAGE and BEST
物理学;天文学
Kazalov, V.V.^1 ; Gavrin, V.N.^1 ; Gorbachev, V.V.^1 ; Gavriljuk, Yu.M.^1 ; Ibragimova, T.V.^1 ; Kalikhov, A.V.^1 ; Shikhin, A.A.^1
Institute for Nuclear Research, Russian Academy of Sciences, prospekt 60-letiya Oktyabrya 7a, Moscow
117312, Russia^1
关键词: Anti-coincidence detectors;    Anticoincidence shield;    Complex shapes;    Detector response functions;    Registration efficiencies;    Registration systems;    Response functions;   
Others  :  https://iopscience.iop.org/article/10.1088/1742-6596/798/1/012200/pdf
DOI  :  10.1088/1742-6596/798/1/012200
学科分类:天文学(综合)
来源: IOP
PDF
【 摘 要 】

Response function simulation using Geant 4 for the detector based on NaI crystal of complex shape in registration systems for the SAGE and BEST experiments is presented. Cylindric NaI crystal has a large well for placing up to eight proportional counters. The detector is using as anti-coincidence shield for counters and an instrument for analysis of different γ-rays sources. The result of detector response function simulation for different background sources and their registration efficiency are given.