Applications of Novel Scintillators for Research and Industry
Phototube non-linearity correction technique
物理学;工业技术
Riboldi, S.^1,2 ; Blasi, N.^1 ; Brambilla, S.^1 ; Camera, F.^1,2 ; Giaz, A.^1 ; Million, B.^1
Istituto Nazionale di Fisica Nucleare INFN, Sezione di Milano, Via Celoria 16, Milan (Mi)
20133, Italy^1
Università degli Studi di Milano, Via Celoria 16, Milan (Mi)
20133, Italy^2
关键词: Automatic procedures;    Calibration information;    Calibration source;    Linearity corrections;    Non-linear devices;    Photomultiplier tube;    Scintillation light;    Signal/noise ratio;   
Others  :  https://iopscience.iop.org/article/10.1088/1742-6596/620/1/012007/pdf
DOI  :  10.1088/1742-6596/620/1/012007
学科分类:工业工程学
来源: IOP
PDF
【 摘 要 】

Scintillation light is often detected by photo-multiplier tube (PMT) technology. PMTs are however intrinsically non linear devices, especially when operated with high light yield scintillators and high input photon flux. Many physical effects (e.g. inter-dynode field variation, photocathode resistivity, etc.) can spoil the ideal PMT behavior in terms of gain, ending up in what are addressed as the under-linearity and over-linearity effects. Established techniques implemented in the PMT base (e.g. increasing bleeding current, active voltage divider, etc.) can mitigate these effects, but given the unavoidable spread in manufacturing and materials, it turns out that, with respect to linearity at the percent level, every PMT sample is a story of its own. The residual non linearity is usually accounted for with polynomial correction of the spectrum energy scale, starting from the position of a few known energy peaks of calibration sources, but uncertainly remains in between of calibration peaks. We propose to retrieve the calibration information from the entire energy spectrum and not only the position of full energy peaks (FEP), by means of an automatic procedure that also takes into account the quality (signal/noise ratio) of the information about the non-linearity extracted from the various regions of the spectrum.

【 预 览 】
附件列表
Files Size Format View
Phototube non-linearity correction technique 1216KB PDF download
  文献评价指标  
  下载次数:14次 浏览次数:41次