17th International Conference on the Use of Computers in Radiation Therapy | |
Dose distribution transfer from CyberKnife to Varian treatment planning system | |
物理学;计算机科学 | |
Osewski, W.^1 ; S´losarek, K.^2 ; Karaszewska, B.^2 | |
Radiotherapy Department, Maria Sklodowska-Curie Memorial Cancer Center, Institute of Oncology, Wybrzee Armii Krajowej 15, 44-101 Gliwice, Poland^1 | |
Radiotherapy and Brachytherapy Planning Department, Maria Sklodowska-Curie Memorial Cancer Center, Institute of Oncology, Wybrzee Armii Krajowej 15, 44-101 Gliwice, Poland^2 | |
关键词: Cyberknife; DICOM format; Different treatments; Dose distributions; Error messages; Medical data; Treatment planning systems; | |
Others : https://iopscience.iop.org/article/10.1088/1742-6596/489/1/012056/pdf DOI : 10.1088/1742-6596/489/1/012056 |
|
学科分类:计算机科学(综合) | |
来源: IOP | |
【 摘 要 】
The aim of this paper was to introduce one of the options of the locally developed DDcon.exe which gives the possibility to transfer the dose distribution from CyberKnife (Accuray) treatment planning system (CK TPS) to Varian treatment planning system (Eclipse TPS, Varian). DICOM format is known as a universal format for medical data. The dose distribution is stored as RTdose file in DICOM format, so there should be a possibility to transfer it between different treatment planning systems. Trying to transfer RTdose file from CK TPS to Eclipse TPS the error message occurs. That's because the RTdose file in CK TPS is connected with Structure-Set-Sequence against Eclipse TPS where it's connected with RT-Plan-Sequence. To make it transferable RTdose file from CK TPS have to be 'disconnected' from Structure-Set-Sequence and 'connected' with RT-Plan-Sequence. This is possible thanks DDcon software which creates new RTdose file by changing proper DICOM tags in original RTdose file. New homemade software gives us an opportunity to transfer dose distribution from CyberKnife TPS to TPS Eclipse. This method opens new possibilities to combine or compare different treatment techniques in Varian TPS.
【 预 览 】
Files | Size | Format | View |
---|---|---|---|
Dose distribution transfer from CyberKnife to Varian treatment planning system | 684KB | download |