会议论文详细信息
3rd International Workshop on New Computational Methods for Inverse Problems
Depth-variant blind restoration with pupil-phase constraints for 3D confocal microscopy
物理学;计算机科学
Hadj, Saima Ben^1 ; Blanc-Féraud, Laure^1 ; Engler, Gilbert^2
Morphème Research group-I3S, UNS, 2000 route des Lucioles, 06903 Sophia Antipolis Cedex, France^1
Unit IBSV, INRA, 06903 Sophia Antipolis, France^2
关键词: Biological specimens;    Blind estimation algorithm;    Confocal laser scanning microscopy;    Convex combinations;    Gerchberg-Saxton algorithm;    Physical constraints;    Refractive index mismatch;    Three dimensional images;   
Others  :  https://iopscience.iop.org/article/10.1088/1742-6596/464/1/012009/pdf
DOI  :  10.1088/1742-6596/464/1/012009
学科分类:计算机科学(综合)
来源: IOP
PDF
【 摘 要 】

Three-dimensional images of confocal laser scanning microscopy suffer from a depth-variant blur, due to refractive index mismatch between the different mediums composing the system as well as the specimen, leading to optical aberrations. Our goal is to develop an image restoration method for 3D confocal microscopy taking into account the blur variation with depth. The difficulty is that optical aberrations depend on the refractive index of the biological specimen. The depth-variant blur function or the Point Spread Function (PSF) is thus different for each observation. A blind or semi-blind restoration method needs to be developed for this system. For that purpose, we use a previously developed algorithm for the joint estimation of the specimen function (original image) and the 3D PSF, the continuously depth-variant PSF is approximated by a convex combination of a set of space-invariant PSFs taken at different depths. We propose to add to that algorithm a pupil-phase constraint for the PSF estimation, given by the the optical instrument geometry. We thus define a blind estimation algorithm by minimizing a regularized criterion in which we integrate the Gerchberg-Saxton algorithm allowing to include these physical constraints. We show the efficiency of this method relying on some numerical tests.

【 预 览 】
附件列表
Files Size Format View
Depth-variant blind restoration with pupil-phase constraints for 3D confocal microscopy 564KB PDF download
  文献评价指标  
  下载次数:12次 浏览次数:34次