会议论文详细信息
7th International Conference on 3D Radiation Dosimetry
A potential modification of the γ-evaluation: mapping dose disagreements using γ-vector fields
Holmes, O.^2 ; Darko, J.^1 ; Olding, T.^1 ; Schreiner, L.J.^1,2
Cancer Centre of South Eastern Ontario, Kingston General Hospital, 25 King Street West, Kingston, ON K7L5P9, Canada^1
Department of Physics, Queen's University, Kingston, ON K7L3N6, Canada^2
关键词: Electronic portal imaging devices;    Gel measurement;    Treatment plans;    Vector fields;   
Others  :  https://iopscience.iop.org/article/10.1088/1742-6596/444/1/012085/pdf
DOI  :  10.1088/1742-6596/444/1/012085
来源: IOP
PDF
【 摘 要 】

This work describes the use of a motion phantom and 1D, 2D, and 3D ion chamber, EBT3 film, electronic portal imaging device (EPID) and FXG gel measurements for dosimetric validation of a stereotactic ablative radiation therapy (SBRT) technique in our clinic. Results show good agreement between the measurements and calculated treatment plan dose.

【 预 览 】
附件列表
Files Size Format View
A potential modification of the γ-evaluation: mapping dose disagreements using γ-vector fields 847KB PDF download
  文献评价指标  
  下载次数:29次 浏览次数:39次