会议论文详细信息
7th International Conference on 3D Radiation Dosimetry | |
Preliminary characterization of PRESAGE? for 3D dosimetry of 62 MeV proton beam | |
Gorjiara, T.^1 ; Kacperek, A.^2 ; Kuncic, Z.^1 ; Baldock, C.^1,4 ; Doran, S.^3 | |
Institute of Medical Physics, School of Physics, University of Sydney, NSW 2006, Australia^1 | |
Douglas Cyclotron, Clatterbridge Centre for Oncology, Wirral, United Kingdom^2 | |
CRUK Clinical Magnetic Resonance Research Group, Institute of Cancer Research, Surrey, United Kingdom^3 | |
Faculty of Science, Macquarie University, NSW 2109, Australia^4 | |
关键词: 3d dosimeters; Dose range; Dose response; Heavy particles; Optical absorbance; | |
Others : https://iopscience.iop.org/article/10.1088/1742-6596/444/1/012058/pdf DOI : 10.1088/1742-6596/444/1/012058 |
|
来源: IOP | |
【 摘 要 】
PRESAGE® has previously shown potential for 3D dosimetry of heavy particles. A new formulation has specifically developed for dosimetry of protons/heavy ions. This work provides a preliminary characterization the new formulation of PRESAGE® by measuring optical absorbance and dose response after irradiating by a 62 MeV proton beam for a dose range of 0.5 - 20 Gy. Results show linear dose response and the evolution over time of the optical density of the 3D dosimeter.
【 预 览 】
Files | Size | Format | View |
---|---|---|---|
Preliminary characterization of PRESAGE? for 3D dosimetry of 62 MeV proton beam | 534KB | download |