会议论文详细信息
35th International Symposium on Remote Sensing of Environment
Laboratory geometric calibration of areal digital aerial camera
地球科学;生态环境科学
Yuan, F.^1 ; Qi, W.J.^1 ; Fang, A.P.^1
National Calibration Center for Electro-optical Distance Meter, Chinese Academy of Surveying and Mapping, Beijing 100039, China^1
关键词: Calibration method;    Digital aerial camera;    Distortion parameters;    Geometric calibrations;    High-precision;    Interior orientation;    Non-metric cameras;    Parallel light;   
Others  :  https://iopscience.iop.org/article/10.1088/1755-1315/17/1/012196/pdf
DOI  :  10.1088/1755-1315/17/1/012196
学科分类:环境科学(综合)
来源: IOP
PDF
【 摘 要 】

Digital aerial camera is non-metric camera. Geometric calibration, including the determination of interior orientation elements and distortion parameters, is the base of high precision photogrammetry. In this paper, a laboratory geometric calibration system of areal digital aerial cameras is developed. This system uses a collimator and a star tester as the target generator. After measurement of the coordinates of targets on the CCD plane and corresponding angles of parallel lights, the geometric calibration of digital aerial camera can be realized according to the geometric calibration model of this paper. Geometric calibration experiments are taken out based on this system using two kinds of mainstream digital aerial cameras, Cannon EOS 5D Mark II and Hasselblad H3D. Experiment results show that this system can satisfy the calibration requirements of aerial photogrammetric application and prove the correctness and the reliability of this calibration method.

【 预 览 】
附件列表
Files Size Format View
Laboratory geometric calibration of areal digital aerial camera 357KB PDF download
  文献评价指标  
  下载次数:20次 浏览次数:31次