科技报告详细信息
Performance of a MEMS-Based AO-OCT System Using Fourier Reconstruction.
Evans, J. ; Zawadzki, R. ; Jones, S. ; Olivier, S. ; Werner, J. S.
Technical Information Center Oak Ridge Tennessee
关键词: Adaptive optics;    Optical properties;    Tomography;    Images;    Coherence;   
RP-ID  :  DE2009947762
学科分类:工程和技术(综合)
美国|英语
来源: National Technical Reports Library
PDF
【 摘 要 】

Adaptive optics (AO) and optical coherence tomography (OCT) are powerful imaging modalities that, when combined, can provide high-resolution (3.5 im isotropic), 3-D images of the retina. The AO-OCT system at UC Davis has demonstrated the utility of this technology for microscopic, volumetric, in vivo retinal imaging. The current system uses an AOptix bimorph deformable mirror (DM) for low-order, high-stroke correction and a 140-actuator Boston Micromachines DM for high-order correction. Developments to improve performance or functionality of the instrument are on-going. Based on previous work in system characterization we have focused on improved AO control. We present preliminary results and remaining challenges for a newly implemented Fourier transform reconstructor (FTR). The previously reported error budget analysis is also reviewed and updated, with consideration of how to improve both the amount of residual error and the robustness of the system. Careful characterization of the AO system will lead to improved performance and inform the design of future systems.

【 预 览 】
附件列表
Files Size Format View
DE2009947762.pdf 2097KB PDF download
  文献评价指标  
  下载次数:25次 浏览次数:52次