科技报告详细信息
Calibration Report for the Plutonium Finishing Plant Segmented Gamma Scan Assay System.
Fazzari, D. M.
Technical Information Center Oak Ridge Tennessee
关键词: Calibration;    Validation;    Plutonium;    WIPP;    Measurements;   
RP-ID  :  DE2003806000
学科分类:工程和技术(综合)
美国|英语
来源: National Technical Reports Library
PDF
【 摘 要 】

This document presents the results of the calibration of the Segmented Gamma Scan Assay System (SGSAS) at the Plutonium Finishing Plant (PFP). The calibration was performed in July, 2000 to qualify the system for assay of residues cans from PFP which will be packaged into pipe overpack containers for shipment to WIPP. The system was calibrated using NIST traceable fission product sources and initial calibration verification runs were performed using the same sources. Validation measurements for WIPP QAOs and the safeguards organization also were performed using NIST traceable plutonium standards.

【 预 览 】
附件列表
Files Size Format View
DE2003806000.pdf 1076KB PDF download
  文献评价指标  
  下载次数:11次 浏览次数:16次