This document presents the results of the calibration of the Segmented Gamma Scan Assay System (SGSAS) at the Plutonium Finishing Plant (PFP). The calibration was performed in July, 2000 to qualify the system for assay of residues cans from PFP which will be packaged into pipe overpack containers for shipment to WIPP. The system was calibrated using NIST traceable fission product sources and initial calibration verification runs were performed using the same sources. Validation measurements for WIPP QAOs and the safeguards organization also were performed using NIST traceable plutonium standards.