科技报告详细信息
Multi-Channel Gel Electrophoresis and Continuous Fraction Collection Apparatus for High Throughput Protein Separation and Characterization.
Choi, M. ; Nordmeyer, R. A. ; Biggin, M. D. ; Jin, J.
Technical Information Center Oak Ridge Tennessee
关键词: Proteins;    Separation;    Electrophoresis;    Gels;    Elution;   
RP-ID  :  DE2010983162
学科分类:工程和技术(综合)
美国|英语
来源: National Technical Reports Library
PDF
【 摘 要 】

To facilitate a direct interface between protein separation by PAGE and protein identification by mass spectrometry, we developed a multichannel system that continuously collects fractions as protein bands migrate off the bottom of gel electrophoresis columns. The device was constructed using several short linear gel columns, each of a different percent acrylamide, to achieve a separation power similar to that of a long gradient gel. A Counter Free-Flow elution technique then allows continuous and simultaneous fraction collection from multiple channels at low cost. We demonstrate that rapid, high-resolution separation of a complex protein mixture can be achieved on this system using SDS-PAGE. In a 2.5 h electrophoresis run, for example, each sample was separated and eluted into 48-96 fractions over a mass range of 10-150 kDa; sample recovery rates were 50percent or higher; each channel was loaded with up to 0.3 mg of protein in 0.4 mL; and a purified band was eluted in two to three fractions (200 L/fraction). Similar results were obtained when running native gel electrophoresis, but protein aggregation limited the loading capacity to about 50 g per channel and reduced resolution.

【 预 览 】
附件列表
Files Size Format View
DE2010983162.pdf 700KB PDF download
  文献评价指标  