科技报告详细信息
Sensitivity of Spectroradio-metric Calibrations in the Near Infrared to Variations in Atmospheric Water Vapor.
Meyers, D. R. ; Andreas, A. A.
Technical Information Center Oak Ridge Tennessee
关键词: Spectroradiometers;    Calibration;    Water vapor;    Research;    Photovoltaics;   
RP-ID  :  DE200415007168
学科分类:工程和技术(综合)
美国|英语
来源: National Technical Reports Library
PDF
【 摘 要 】

Spectra of natural sunlight and artificial sources are important in photovoltaic research. Calibration of the spectroradiometers used for these measurements is derived from the response to spectral irradiance standards from the National Institute of Standards and Technology (NIST). Some photovoltaic devices respond in the near infrared, or NIR, so spectral measurements and calibrations are needed in this region. Over the course of several calibrations, we identified variations > 5% in spectroradiometer NIR calibration data for a certain spectroradiometer. A detailed uncertainty analysis did not reflect the observed variation. Reviewing calibration procedures and historical data, we noted that the variations were seen in water vapor absorption bands. We used spectral transmission models to compute changes in atmospheric transmission (as a function of water vapor content) over path lengths occurring during calibration. The results indicate that the observed variations result from varying water vapor content. A correction algorithm for adjusting measured data was developed based on our analysis.

【 预 览 】
附件列表
Files Size Format View
DE200415007168.pdf 583KB PDF download
  文献评价指标  