科技报告详细信息
Response of LaBr(sub 3)(Ce) Detector to 2-11 MeV Gamma Rays.
Seabury, E. H. ; Wharton, J. C. ; Caffrey, A. J.
Technical Information Center Oak Ridge Tennessee
关键词: Scintillation detectors;    Cerium;    Gamma detectors;    Radiation detectors;    Lanthanum bromide;   
RP-ID  :  DE2007911790
学科分类:工程和技术(综合)
美国|英语
来源: National Technical Reports Library
PDF
【 摘 要 】

The development of lanthanum halide scintillation detectors has great potential application in field-portable prompt-gamma neutron activation analysis systems. Because the low-energy response of these detectors has already been well-characterized (1(-(2))), we have measured their response to higher energy gamma rays in the region between 2 and 11 MeV. We have measured the response of a 2-inch (5.08 cm) by 2-inch long LaBr3(Ce) detector to high energy gamma rays produced by neutron interactions on chlorine, hydrogen, iron, nitrogen, phosphorous, and sulfur. The response of the LaBr3(Ce) detector is compared to that of HPGe and NaI(Tl) detectors.

【 预 览 】
附件列表
Files Size Format View
DE2007911790.pdf 311KB PDF download
  文献评价指标  
  下载次数:92次 浏览次数:29次