RHIC Beam Loss Monitor System Initial Operation.
Witkover, R. L. ; Michnoff, R. J. ; Geller, J. M.
Technical Information Center Oak Ridge Tennessee
关键词: Accelerators;    Monitors;    Quenching;    Superconducting magnets;   
RP-ID  :  DE2004770911
学科分类:工程和技术(综合)
美国|英语
来源: National Technical Reports Library
PDF
【 摘 要 】

The RHIC Beam Loss Monitor (BLM) System is designed to prevent beam loss quenching of the superconducting magnets, and acquire loss data. Four hundred ion chambers are located around the rings to detect losses. The required 8-decade range in signal current is compressed using an RC pre-integrator ahead of a low current amplifier. A beam abort may be triggered if fast or slow losses exceed programmable threshold levels. A micro-controller based VME module sets references and gains and reads trip status for up to 64 channels. Results obtained with the detectors in the RHIC Sextant Test and the prototype electronics in the AGS-to-RHIC (AtR) transfer line are presented along with the present status of the system.