科技报告详细信息
Beta and Gamma Correction Factors for the Eberline R0-20 Ionization Chamber Survey Instrument.
Johnson, M. L. ; Rathbone, B. A. ; Bratvold, T.
Technical Information Center Oak Ridge Tennessee
关键词: Isotopes;    Beta particle detectors;    Gamma radiation;    Surveys;    Exposure;   
RP-ID  :  DE2001786805
学科分类:工程和技术(综合)
美国|英语
来源: National Technical Reports Library
PDF
【 摘 要 】
This technical document provides details of derived correction factors for the Eberline R0-20 survey meter, which uses an ionization chamber to measure ambient exposure rates. A thin end window allows the instrument to measure exposure rates from non-penetrating radiation (i.e., beta radiation). Correction factors are provided for contact measurements with beta and gamma disk sources, gamma beams and, finally, general area beta fields. Beta correction factors are based on the instrument's response to 204Tl, selected as the most conservative isotope for beta correction factors, as indicated in previous studies of similar instruments using 204Tl, 147Pm, and 90Sr(Y) isotopes (LANL 1982). Gamma correction factors are based on 137Cs, considered the predominant source of gamma radiation on the Hanford Site.
【 预 览 】
附件列表
Files Size Format View
DE2001786805.pdf 239KB PDF download