科技报告详细信息
Radiation Dose Measurements of the Insertion Devices Using Radiachromic Film Dosimeters.
Alderman, J. ; Semones, E. ; Job, P. K.
Technical Information Center Oak Ridge Tennessee
关键词: Advanced photon source;    Radiation doses;    Dosimeters;    Dosimetry;    Dosemeters;   
RP-ID  :  DE2004834029
学科分类:工程和技术(综合)
美国|英语
来源: National Technical Reports Library
PDF
【 摘 要 】

The Advanced Photon Source (APS) uses Nd-Fe-B permanent magnets in the insertion devices to produce x-rays for scientific research. Earlier investigations have exhibited varying degrees of demagnetization of these magnets due to irradiation from electron beams, (sup 60)Co (gamma)-rays, and high-energy neutrons. Radiation-induced demagnetization has been observed in the APS insertion devices and was first measured in December of 2001. Partial demagnetization has also been observed in insertion devices at the European Synchrotron Radiation Facility (ESRF), where Nd-Fe-B permanent magnets are also used. Growing concern for the lifetime of APS insertion devices, as well as the permanent magnets that will be used in next-generation, high-power light sources, like the FEL, resulted from the partial demagnetization observations made at both facilities. This concern in relation to radiation-induced demagnetization spurred a long-term project to measure and analyze the absorbed doses received by the APS insertion devices. The project required a reliable photon high-dose dosimetry technique capable of measuring absorbed doses greater than 10(sup 6) rad, which was not readily available at the APS. Through a collaboration with the National Institute of Standards and Technology (NIST), one such technique using radiachromic films was considered, tested, and calibrated at the APS. This consequently led to the implementation of radiachromic film dosimetry for measuring the absorbed doses received by the insertion devices for each of the APS runs.

【 预 览 】
附件列表
Files Size Format View
DE2004834029.pdf 180KB PDF download
  文献评价指标  
  下载次数:14次 浏览次数:36次