科技报告详细信息
High Bias of DCAP and Aa Analyses of Hg(I).
Holcomb, H. P.
Technical Information Center Oak Ridge Tennessee
关键词: Bias;    Spectrometry;    Mercury;    Absorption spectroscopy;    Performance;   
RP-ID  :  DE2001781331
学科分类:工程和技术(综合)
美国|英语
来源: National Technical Reports Library
PDF
【 摘 要 】

This brief, but definitive, study indicates that the relatively new spectrometric techniques, DCAP and AA (Atomic Absorption), give high values for mercury when employed to analyze solutions that contain Hg(I) per se or that contain soluble mercury which is or has been subject to reducing conditions.

【 预 览 】
附件列表
Files Size Format View
DE2001781331.pdf 173KB PDF download