科技报告详细信息
High Bias of DCAP and Aa Analyses of Hg(I). | |
Holcomb, H. P. | |
Technical Information Center Oak Ridge Tennessee | |
关键词: Bias; Spectrometry; Mercury; Absorption spectroscopy; Performance; | |
RP-ID : DE2001781331 | |
学科分类:工程和技术(综合) | |
美国|英语 | |
来源: National Technical Reports Library | |
【 摘 要 】
This brief, but definitive, study indicates that the relatively new spectrometric techniques, DCAP and AA (Atomic Absorption), give high values for mercury when employed to analyze solutions that contain Hg(I) per se or that contain soluble mercury which is or has been subject to reducing conditions.