科技报告详细信息
SXR Continuum Radiation Transmitted Through Metallic Filters: An Analytical Approach To Fast Electron Temperature Measurements
Delgado-Aparicio, L. ; Tritz, K. ; Kramer, T. ; Stutman, D. ; Finkentha, M. ; Hill, K. ; Bitter, M.
Princeton University. Plasma Physics Laboratory.
关键词: Plasma X-Ray, Srx;    X-Ray, Srx;    70 Plasma Physics And Fusion Technology;    Thermonuclear Devices;    Soft X Radiation;   
DOI  :  10.2172/988882
RP-ID  :  PPPL-4543
RP-ID  :  DE-ACO2-09CH11466
RP-ID  :  988882
美国|英语
来源: UNT Digital Library
PDF
【 摘 要 】

A new set of analytic formulae describes the transmission of soft X-ray (SXR) continuum radiation through a metallic foil for its application to fast electron temperature measurements in fusion plasmas. This novel approach shows good agreement with numerical calculations over a wide range of plasma temperatures in contrast with the solutions obtained when using a transmission approximated by a single-Heaviside function [S. von Goeler, Rev. Sci. Instrum., 20, 599, (1999)]. The new analytic formulae can improve the interpretation of the experimental results and thus contribute in obtaining fast teperature measurements in between intermittent Thomson Scattering data.

【 预 览 】
附件列表
Files Size Format View
988882.pdf 630KB PDF download
  文献评价指标  