科技报告详细信息
Evaluation Of Automated Low-Field Nuclear Magnetic Resonance (NMR) Relaxometry For Analysis Of Silicone Polymers
Wilson, M. H.
Kansas City Plant (U.S.)
关键词: Polymers;    36 Materials Science;    Nuclear Magnetic Resonance;    Service Life;    Silicones;   
DOI  :  10.2172/972918
RP-ID  :  KCP-613-8630s
RP-ID  :  DE-AC04-01AL66850
RP-ID  :  972918
美国|英语
来源: UNT Digital Library
PDF
【 摘 要 】
Screening studies and Design of Experiments (DoE) were performed to evaluate measurement variation of a new, non-destructive Nuclear Magnetic Resonance (NMR) test system designed to assess age-induced degradation of Outer Pressure Pads (OPP). The test method and results from 54,275 measurements are described. A reduction in measurement error was obtained after metal support struts were replaced with plastic support struts adjacent to the front position of the test chamber. However, remaining interference and a lack of detecting any age-related degradation prevent the use of the NMR system as a non-destructive surveillance test for OPPs. A cursory evaluation of the system with cellular silicone samples obtained more uniform results with increased error as measurements approached the sample’s edge.
【 预 览 】
附件列表
Files Size Format View
972918.pdf 716KB PDF download
  文献评价指标  
  下载次数:11次 浏览次数:76次