| Evaluation of Automated Low-Field NMR Relaxometry For Analysis of Polymers | |
| Wilson, Mark | |
| Kansas City Plant (KCP), Kansas City, MO | |
| 关键词: Polymers; 37 Inorganic, Organic, Physical And Analytical Chemistry; Test Facilities; Performance; Nuclear Magnetic Resonance; | |
| DOI : 10.2172/952481 RP-ID : KCP-613-8531 RP-ID : DE-AC04-01AL66850 RP-ID : 952481 |
|
| 美国|英语 | |
| 来源: UNT Digital Library | |
PDF
|
|
【 摘 要 】
Screening studies and a Design of Experiment were performed to evaluate measurement variation of a new, non-destructive NMR test system designed to assess age induced degradation of Outer Pressure Pads. The test method and results from 76,746 measurements are described indicating that a metal support strut adjacent to the front position of the test chamber interferes with the measurements and that there is a measurement property gradient from the front to the back of the chamber. Also, obvious compression set areas on a surveillance OPP were not detected, but hidden, internal voids within a newly molded OPP were detected.
【 预 览 】
| Files | Size | Format | View |
|---|---|---|---|
| 952481.pdf | 521KB |
PDF