| Further improvement in the precision of 233U measurements by accelerator mass spectrometry at the Lawrence Livermore National Laboratory. | |
| Tumey, S. J. ; Brown, T. A. ; Hamilton, T. F. ; Buchholz, B. A. | |
| Lawrence Livermore National Laboratory | |
| 关键词: Ionization Chambers; Modifications; Lawrence Livermore National Laboratory; 46 Instrumentation Related To Nuclear Science And Technology; Faraday Cups; | |
| DOI : 10.2172/926045 RP-ID : LLNL-TR-401375 RP-ID : W-7405-ENG-48 RP-ID : 926045 |
|
| 美国|英语 | |
| 来源: UNT Digital Library | |
PDF
|
|
【 摘 要 】
In response to sponsor interest in October 2005 we proposed two methods for enhancing the precision of {sup 233}U accelerator mass spectrometry (AMS) capabilities at the Lawrence Livermore National Laboratory (LLNL). In a previous report we evaluated the first method and demonstrated that by increasing sample loading by a factor of four coupled with performing four replicates of each sample, we could achieve measurement precision of {approx}1%. Recent modifications to our system have enabled us to test the second proposed method. By changing our setup to normalize {sup 233}U ions counted in a gas ionization chamber to {sup 238}U measured as a current in an off-axis Faraday cup we were able to attain 1% precision without the need for replicate analysis. This method could be further refined to achieve 0.5% precision in samples of interest.
【 预 览 】
| Files | Size | Format | View |
|---|---|---|---|
| 926045.pdf | 85KB |
PDF