Electronics Personal Dosemeter (EPD-N) Test and Evaluation Report | |
Shourbaji, A.A. | |
Oak Ridge National Laboratory | |
关键词: Calibration; Manufacturers; Photons; Dosimetry; Dosemeters; | |
DOI : 10.2172/885765 RP-ID : ORNL/TM-2002/172 RP-ID : DE-AC05-00OR22725 RP-ID : 885765 |
|
美国|英语 | |
来源: UNT Digital Library | |
【 摘 要 】
Three electronic personal dosimeters (EPD-N) manufactured by Siemens, serial numbers 0635, 0658, and 0683, were tested at the Radiation Calibration Laboratory for an evaluation of their response to neutron, gamma and x-ray radiation. Designed to provide real-time neutron and photon dosimetry, the EPD-N is capable of estimating and displaying neutron and gamma dose components separately for a range of energies from 50 keV to 7 MeV for photon and 0.025 eV to > 10 MeV for neutron. All tests were conducted using the factory calibrations. A technical representative of the manufacturer indicated that site-specific calibrations are required as factory settings are calibrated for the lowest neutron energy limit of 0.025 eV. This raises concerns about the reliability of these devices in measuring neutrons when calibrations are made for a specific site radiological characterization then used at another site.【 预 览 】