科技报告详细信息
Comparison of STIS and SNAP spectrograph throughputs | |
Aldering, Greg | |
Lawrence Berkeley National Laboratory | |
关键词: Spectrometers; Comparative Evaluations; Performance; 47 Other Instrumentation; | |
DOI : 10.2172/842491 RP-ID : LBNL--51156 RP-ID : AC03-76SF00098 RP-ID : 842491 |
|
美国|英语 | |
来源: UNT Digital Library | |
【 摘 要 】
This is a comparison of the measured throughput of STIS on HST versus what we might expect from the spectrograph on SNAP. The principle reference is Woodgate et al. (1998) PASP, 110, 1183. Additional material was taken from the STIS Handbook, available on-line at www.stsci.edu. The goal is to demonstrate that there are sound reasons to expect better performance for a SNAP spectrograph (even one with a grating) than would be expected by scaling from HST+STIS.
【 预 览 】
Files | Size | Format | View |
---|---|---|---|
842491.pdf | 35KB | download |