科技报告详细信息
MESERAN Calibration for Low Level Organic Residues
Benkovich, M.G.
Kansas City Plant (U.S.)
关键词: Calibration;    46 Instrumentation Related To Nuclear Science And Technology;    Monitoring;    Organic Compounds;    Geiger-Mueller Counters;   
DOI  :  10.2172/822704
RP-ID  :  KCP-613-6792
RP-ID  :  ACO4-01AL66850
RP-ID  :  822704
美国|英语
来源: UNT Digital Library
PDF
【 摘 要 】

Precision cleaning studies done at Honeywell Federal Manufacturing & Technologies (FM&T), the Kansas City Plant (KCP), and at other locations within the Department of Energy (DOE) Weapons complex over the last 30 years have depended upon results from MESERAN Evaporative Rate Analysis for detecting low levels of organic contamination. The characterization of the surface being analyzed is carried out by depositing a Carbon-14 tagged radiochemical onto the test surface and monitoring the rate at which the radiochemical disappears from the surface with a Geiger-Mueller counter. In the past, the total number of counts over a 2-minute span have been used to judge whether a surface is contaminated or not and semi-quantitatively to what extent. This technique is very sensitive but has not enjoyed the broad acceptance of a purely quantitative analysis. The work on this project developed calibrations of various organic contaminants typically encountered in KCP operations. In addition, a new analysis method was developed to enhance the ability of MESERAN Analyzers to detect organic contamination and yield quantitative data in the microgram and nanogram levels.

【 预 览 】
附件列表
Files Size Format View
822704.pdf 1225KB PDF download
  文献评价指标  
  下载次数:1次 浏览次数:6次