【 摘 要 】

An Eberline continuous alpha air monitor model Alpha 7L was evaluated at Oak Ridge National Laboratory (ORNL) using the capabilities available at the Environmental Effects Laboratory (EELab). A series of tests were performed to ensure that procured units meet the requirements of the purchasing facility, Los Alamos National Laboratory (LANL). In addition to reporting on the results of each test, other activities were performed to reduce discovered susceptibilities. The parameters monitored during the tests typically included the airflow rate and/or net {sup 239}Pu concentration values (pCi/liter). In addition, the spectrum display and operational status were monitored. Follow up tests were also performed on two LANL-provided production units. The results of those tests are at the end of this report.