| Dewarless Logging Tool - 1st Generation | |
| HENFLING,JOSEPH A. ; NORMANN,RANDY A. | |
| Sandia National Laboratories | |
| 关键词: Geothermal Legacy; Thermal Insulation; Reservoir Temperature Geothermal Legacy; Geothermal Wells; 15 Geothermal Energy; | |
| DOI : 10.2172/763144 RP-ID : SAND2000-1680 RP-ID : AC04-94AL85000 RP-ID : 763144 |
|
| 美国|英语 | |
| 来源: UNT Digital Library | |
PDF
|
|
【 摘 要 】
This report focuses on Sandia National Laboratories' effort to create high-temperature logging tools for geothermal applications without the need for heat shielding. One of the mechanisms for failure in conventional downhole tools is temperature. They can only survive a limited number of hours in high temperature environments. For the first time since the evolution of integrated circuits, components are now commercially available that are qualified to 225 C with many continuing to work up to 300 C. These components are primarily based on Silicon-On-Insulator (SOI) technology. Sandia has developed and tested a simple data logger based on this technology that operates up to 300 C with a few limiting components operating to only 250 C without thermal protection. An actual well log to 240 C without shielding is discussed. The first prototype high-temperature tool measures pressure and temperature using a wire-line for power and communication. The tool is based around the HT83C51 microcontroller. A brief discussion of the background and status of the High Temperature Instrumentation program at Sandia, objectives, data logger development, and future project plans are given.
【 预 览 】
| Files | Size | Format | View |
|---|---|---|---|
| 763144.pdf | 1085KB |
PDF