科技报告详细信息
Neutron Coincidence Counting Studies | |
Rogers, Jeremy L. ; Ely, James H. ; Kouzes, Richard T. ; Lintereur, Azaree T. ; Siciliano, Edward R. | |
Pacific Northwest National Laboratory (U.S.) | |
关键词: Boron-10; He3 Alternative; Coincidence Counter; Safeguards; | |
DOI : 10.2172/1051994 RP-ID : PNNL-21686 RP-ID : AC05-76RL01830 RP-ID : 1051994 |
|
美国|英语 | |
来源: UNT Digital Library | |
【 摘 要 】
The efficiency comparison for measured and simulated responses of a 10B-lined proportional counter and a 3He proportional counter in a close, symmetrical geometry are presented. The measurement geometry was modeled in MCNPX to validate the methods used for simulating the response of both the 3He and 10B-lined tubes. The MCNPX models agree within 1% with the 3He tube measurements and within 3% for the 10B-lined tubes when a 0.75-µm boron-metal lining is used.
【 预 览 】
Files | Size | Format | View |
---|---|---|---|
1051994.pdf | 2745KB | download |