【 摘 要 】

ORNL's Technical Testing and Analysis Center (TTAC) acquired a High Purity Germanium Detector (HPGe) from ORTEC - a variant called an Interchangeable Detection Module (IDM). This detector has excellent energy resolution as well as high intrinsic efficiency. The purpose of this report is to detail the determination of the efficiency curve of the IDM, so future measurements can quantify the (otherwise unknown) activity of sources. Without such a curve, the activity cannot be directly reported by use of the IDM alone - a separate device such as an ion chamber would be required. This builds upon the capability of TTAC. The method for determining the energy-dependent intrinsic efficiency is laid-out in this report. It's noteworthy that this basic technique can be applied to any spectroscopic radiation detector, independent of the specific type (e.g. NaI, CzT, ClYC).

【 预 览 】
附件列表
Files Size Format View
1048166.pdf 848KB PDF download
  文献评价指标  
  下载次数:20次 浏览次数:63次