科技报告详细信息
Calibration and Use of the Canberra iSolo 300G
Smith, T ; Graham, C L ; Sundsmo, T ; Shingleton, K L
Lawrence Livermore National Laboratory
关键词: Calibration;    61 Radiation Protection And Docimetry;    Air Filters;    Radon;    Radioisotopes;   
DOI  :  10.2172/1017997
RP-ID  :  LLNL-TR-463319
RP-ID  :  W-7405-ENG-48
RP-ID  :  1017997
美国|英语
来源: UNT Digital Library
PDF
【 摘 要 】

This procedure provides instructions for the calibration and use of the Canberra iSolo Low Background Alpha/Beta Counting System (iSolo) that is used for counting air filters and swipe samples. This detector is capable of providing radioisotope identification (e.g., it can discriminate between radon daughters and plutonium). This procedure includes step-by-step instructions for: (1) Performing periodic or daily 'Background' and 'Efficiency QC' checks; (2) Setting-up the iSolo for counting swipes and air filters; (3) Counting swipes and air filters for alpha and beta activity; and (4) Annual calibration.

【 预 览 】
附件列表
Files Size Format View
1017997.pdf 406KB PDF download
  文献评价指标  