科技报告详细信息
Tests of Coordinate Transfer from Magnetic to Mechanical Reference for LCLS Undulator Fiducialization
Levashov, Yu.
SLAC National Accelerator Laboratory
关键词: Calibration;    Light Sources;    Measuring Methods;    43 Particle Accelerators;    Performance Xfel;   
DOI  :  10.2172/1000390
RP-ID  :  SLAC-TN-10-088
RP-ID  :  AC02-76SF00515
RP-ID  :  1000390
美国|英语
来源: UNT Digital Library
PDF
【 摘 要 】

Fiducialization of the LCLS undulators will be based on magnetic measurements by Hall probe. Pointed magnets, proposed by I.Vasserman for quadrupole lens fiducialization will be used as an intermediate reference. A prototype of the pointed magnet fixture has been made and tested. In this note we will describe a procedure for measuring the position of the center of the Hall probe sensitive area with respect to the undulator fiducial marks. The pointed magnet calibration procedure, a two-point algorithm for locating the magnetic center of the fixture, and test results are presented.

【 预 览 】
附件列表
Files Size Format View
1000390.pdf 1454KB PDF download
  文献评价指标  
  下载次数:20次 浏览次数:51次