科技报告详细信息
Terahertz detectors for long wavelength multi-spectral imaging.
Lyo, Sungkwun Kenneth ; Wanke, Michael Clement ; Reno, John Louis ; Shaner, Eric Arthur ; Grine, Albert D.
关键词: FIELD EFFECT TRANSISTORS;    PLASMONS;    SENSITIVITY;    SPECTROSCOPY;    WAVELENGTHS Spectral analysis-Instruments;    Spectrometer.;    Plasmons (Physics);    Spectroscopy.;    Spectral data.;   
DOI  :  10.2172/921735
RP-ID  :  SAND2007-6354
PID  :  OSTI ID: 921735
Others  :  TRN: US0802157
学科分类:物理(综合)
美国|英语
来源: SciTech Connect
PDF
【 摘 要 】
The purpose of this work was to develop a wavelength tunable detector for Terahertz spectroscopy and imaging. Our approach was to utilize plasmons in the channel of a specially designed field-effect transistor called the grating-gate detector. Grating-gate detectors exhibit narrow-linewidth, broad spectral tunability through application of a gate bias, and no angular dependence in their photoresponse. As such, if suitable sensitivity can be attained, they are viable candidates for Terahertz multi-spectral focal plane arrays. When this work began, grating-gate gate detectors, while having many promising characteristics, had a noise-equivalent power (NEP) of only 10{sup -5} W/{radical}Hz. Over the duration of this project, we have obtained a true NEP of 10{sup -8} W/{radical}Hz and a scaled NEP of 10{sup -9}W/{radical}Hz. The ultimate goal for these detectors is to reach a NEP in the 10{sup -9{yields}-10}W/{radical}Hz range; we have not yet seen a roadblock to continued improvement.
【 预 览 】
附件列表
Files Size Format View
RO201705190000692LZ 1164KB PDF download
  文献评价指标  
  下载次数:3次 浏览次数:29次