科技报告详细信息
EVALUATING EFFECTS OF NEPTUNIUM ON THE SRS METHOD FOR CONTROLLED POTENTIAL COULOMETRIC ASSAY OF PLUTONIUM IN SULFURIC ACID SUPPORTING ELECTROLYTE
Holland, M ; Sheldon Nichols, S
关键词: PLUTONIUM;    VOLTAMETRY;    NEPTUNIUM;    VALENCE;    INTERFERING ELEMENTS;    QUALITY CONTROL;   
DOI  :  10.2172/929466
RP-ID  :  WSRC-STI-2008-00238
PID  :  OSTI ID: 929466
Others  :  TRN: US0806619
学科分类:化学(综合)
美国|英语
来源: SciTech Connect
PDF
【 摘 要 】

A study of the impact of neptunium on the coulometric assay of plutonium in dilute sulfuric acid was performed. Weight aliquots of plutonium standard solutions were spiked with purified neptunium solution to evaluate plutonium measurement performance for aliquots with Pu:Np ratios of 50:1, 30:1, 20:1, 15:1, and 10:1. Weight aliquots of the pure plutonium standard solution were measured as controls. Routine plutonium instrument control standards were also measured. The presence of neptunium in plutonium aliquots significantly increases the random uncertainty associated with the plutonium coulometric measurement performed in accordance with ISO12183:2005.7 However, the presence of neptunium does not appear to degrade electrode performance and conditioning as aliquots of pure plutonium that were interspersed during the measurement of the mixed Pu:Np aliquots continued to achieve the historical short-term random uncertainty for the method. Lack of adequate control of the neptunium oxidation state is suspected to be the primary cause of the elevated measurement uncertainty and will be pursued in a future study.

【 预 览 】
附件列表
Files Size Format View
RO201705180001963LZ 18519KB PDF download
  文献评价指标  
  下载次数:2次 浏览次数:7次