Analytical Method for the Detection of Ozone Depleting Chemicals (ODC) in Commercial Products Using a Gas Chromatograph with an Electron Capture Detector (GC-ECD) | |
Lee, Richard N. ; Dockendorff, Brian P. ; Wright, Bob W. | |
关键词: OZONE LAYER; DAMAGE; POLLUTANTS; TRACE AMOUNTS; DETECTION; ELECTRON CAPTURE; GAS CHROMATOGRAPHY; MEASURING METHODS; ELECTRONIC CIRCUITS; IMPORTS Ozone Depleting Chemicals; GC-ECD; Printed circuit boards; | |
DOI : 10.2172/944011 RP-ID : PNNL-16813 PID : OSTI ID: 944011 Others : Other: 400480000 Others : TRN: US200902%%517 |
|
美国|英语 | |
来源: SciTech Connect | |
【 摘 要 】
This document describes an analytical procedure that was developed for the trace level detection of residual ozone depleting chemicals (ODC) associated with the manufacture of selected commercial products. To ensure the United States meets it obligation under the Montreal Protocol, Congress enacted legislation in 1989 to impose an excise tax on electronic goods imported into the United States that were produced with banned chemicals. This procedure was developed to technically determine if residual ODC chemicals could be detected on electronic circuit boards. The analytical method utilizes a âpurge and trapâ technique followed by gas chromatography with electron capture detection to capture and analyze the volatile chemicals associated with the matrix. The method describes the procedure, the hardware, operating conditions, calibration, and quality control measures in sufficient detail to allow the capability to be replicated. This document corresponds to internal Standard Operating Procedure (SOP) EFL-130A, Rev 4.
【 预 览 】
Files | Size | Format | View |
---|---|---|---|
RO201705180001383LZ | 1086KB | download |