Elemental and Isotopic Analysis of Uranium Oxide an NIST Glass Standards by FEMTOSECOND-LA-ICP-MIC-MS
Ebert, Chris ; Zamzow, Daniel S. ; McBay, Eddie H. ; Bostick, Debra A. ; Bajic, Stanley J. ; Baldwin, David P. ; Houk, R.S.
关键词: ABLATION;    ACCURACY;    ACTINIDE NUCLEI;    CONCENTRATION RATIO;    GLASS;    HEATING;    LASERS;    MONITORING;    NITROCELLULOSE;    PARTICULATES;    PROLIFERATION;    SAMPLE PREPARATION;    SAMPLING;    SENSITIVITY;    SPECTROMETERS;    TRANSIENTS;    URANIUM OXIDES;   
DOI  :  10.2172/963980
RP-ID  :  IS-5194
PID  :  OSTI ID: 963980
Others  :  TRN: US0903451
学科分类:核物理和高能物理
美国|英语
来源: SciTech Connect
PDF
【 摘 要 】
The objective of this work was to test and demonstrate the analytical figures of merit of a femtosecond-laser ablation (fs-LA) system coupled with an inductively coupled plasma-multi-ion collector-mass spectrometer (ICP-MIC-MS). The mobile fs-LA sampling system was designed and assembled at Ames Laboratory and shipped to Oak Ridge National Laboratory (ORNL), where it was integrated with an ICP-MIC-MS. The test period of the integrated systems was February 2-6, 2009. Spatially-resolved analysis of particulate samples is accomplished by 100-shot laser ablation using a fs-pulsewidth laser and monitoring selected isotopes in the resulting ICP-MS transient signal. The capability of performing high sensitivity, spatially resolved, isotopic analyses with high accuracy and precision and with virtually no sample preparation makes fs-LA-ICP-MIC-MS valuable for the measurement of actinide isotopes at low concentrations in very small samples for nonproliferation purposes. Femtosecond-LA has been shown to generate particles from the sample that are more representative of the bulk composition, thereby minimizing weaknesses encountered in previous work using nanosecond-LA (ns-LA). The improvement of fs- over ns-LA sampling arises from the different mechanisms for transfer of energy into the sample in these two laser pulse-length regimes. The shorter duration fs-LA pulses induce less heating and cause less damage to the sample than the longer ns pulses. This results in better stoichiometric sampling (i.e., a closer correlation between the composition of the ablated particles and that of the original solid sample), which improves accuracy for both intra- and inter-elemental analysis. The primary samples analyzed in this work are (a) solid uranium oxide powdered samples having different {sup 235}U to {sup 238}U concentration ratios, and (b) glass reference materials (NIST 610, 612, 614, and 616). Solid uranium oxide samples containing {sup 235}U in depleted, natural, and enriched abundances were analyzed as particle aggregates immobilized in a collodion substrate. The uranium oxide samples were nuclear reference materials (CRMs U0002, U005-A, 129-A, U015, U030-A, and U050) obtained from New Brunswick Laboratory-USDOE.
【 预 览 】
附件列表
Files Size Format View
RO201705170001900LZ 177KB PDF download
  文献评价指标  
  下载次数:16次 浏览次数:34次