WDR-PK-AK-018 | |
Hollister, R | |
关键词: R FACILITIES; CALIBRATION; COLLIMATORS; EFFICIENCY; GEOMETRY; GERMANIUM; RADIOACTIVITY; SEALED SOURCES; SPECTRA; | |
DOI : 10.2172/973642 RP-ID : LLNL-TR-416262 PID : OSTI ID: 973642 Others : TRN: US1002196 |
|
学科分类:核能源与工程 | |
美国|英语 | |
来源: SciTech Connect | |
![]() |
【 摘 要 】
Method - CES SOP-HW-P556 'Field and Bulk Gamma Analysis'. Detector - High-purity germanium, 40% relative efficiency. Calibration - The detector was calibrated on February 8, 2006 using a NIST-traceable sealed source, and the calibration was verified using an independent sealed source. Count Time and Geometry - The sample was counted for 20 minutes at 72 inches from the detector. A lead collimator was used to limit the field-of-view to the region of the sample. The drum was rotated 180 degrees halfway through the count time. Date and Location of Scans - June 1,2006 in Building 235 Room 1136. Spectral Analysis Spectra were analyzed with ORTEC GammaVision software. Matrix and geometry corrections were calculated using OR TEC Isotopic software. A background spectrum was measured at the counting location. No man-made radioactivity was observed in the background. Results were determined from the sample spectra without background subtraction. Minimum detectable activities were calculated by the Nureg 4.16 method. Results - Detected Pu-238, Pu-239, Am-241 and Am-243.【 预 览 】