G-Band Vapor Radiometer Profiler (GVRP) Handbook | |
Caddeau, MP | |
Pacific Northwest National Laboratory | |
关键词: ATMOSPHERIC CHEMISTRY; BRIGHTNESS; GHZ RANGE 100-1000; HUMIDITY; MANUALS; MEASURING INSTRUMENTS; RADIOMETERS; PERFORMANCE; SENSITIVITY; WATER; WATER VAPOR; TECHNOLOGY ASSESSMENT; | |
DOI : 10.2172/982364 RP-ID : DOE/SC-ARM/TR-091 PID : OSTI ID: 982364 Others : TRN: US201020%%166 |
|
美国|英语 | |
来源: SciTech Connect | |
【 摘 要 】
The G-Band Vapor Radiometer Profiler (GVRP) provides time-series measurements of brightness temperatures from 15 channels between 170 and 183.310 GHz. Atmospheric emission in this spectral region is primarily due to water vapor, with some influence from liquid water. Channels between 170.0 and 176.0 GHz are particularly sensitive to the presence of liquid water. The sensitivity to water vapor of the 183.31-GHz line is approximately 30 times higher than at the frequencies of the two-channel microwave radiometer (MWR) for a precipitable water vapor (PWV) amount of less than 2.5 mm. Measurements from the GVRP instrument are therefore especially useful during low-humidity conditions (PWV < 5 mm). In addition to integrated water vapor and liquid water, the GVRP can provide low-resolution vertical profiles of water vapor in very dry conditions.