科技报告详细信息
Qualification of a Carbon Analyzer to Support the Defense Waste Processing Facility
Edwards, T. ; Feller, M.
关键词: CALIBRATION;    CARBON;    EVAPORATORS;    PRODUCTION;    RECOMMENDATIONS;    SLUDGES;    TANKS;    TESTING;    WASTE PROCESSING;   
DOI  :  10.2172/1018309
RP-ID  :  SRNL-STI-2010-00755
PID  :  OSTI ID: 1018309
Others  :  TRN: US201114%%127
学科分类:环境科学(综合)
美国|英语
来源: SciTech Connect
PDF
【 摘 要 】
The I-O Model 1030 carbon analyzer has been qualified for use at the Defense Waste Processing Facility (DWPF). The qualification was a side-by-side comparison of the Model 1030 system with the currently used Model 1010 Analyzer. This recommendation is based on side-by-side comparisons of the new unit to the currently used Model 1010 analyzer that are presented in this report. The side-by-side testing included standards and process samples. The standards, which were used for instrument calibration verifications in the measurement of total inorganic carbon (TIC) and of total organic carbon (TOC), were traceable back to the National Institute of Standards and Technology. The process samples included TIC analyses of Sludge Receipt and Adjustment Tank samples and TOC analyses for Slurry Mix Evaporator (SME) samples. After the Model 1030 has been used for production reporting, DWPF should consider an investigation into the uncertainties associated with the TOC measurements to determine how far below the 18,916 ppm limit DWPF must control the average of the measurements for a set of SME samples to account for the uncertainties of the measurements from this new analyzer. Based upon the results presented in this report, it is recommended that the Model 1030 carbon analyzer is qualified for use. This recommendation is based on side-by-side comparisons of the new unit to the currently used Model 1010 analyzer that are presented in this report. The side-by-side testing included standards for instrument calibration verifications for TIC and TOC, and process samples. The standards were traceable back to NIST. The process samples included TIC analyses of SRAT Receipt samples and TOC analyses for SME samples. At some point in the future, after the Model 1030 has been used for production reporting, DWPF should consider an investigation into the uncertainties associated with the TOC measurements to determine how far below the 18,916 ppm limit DWPF must control the average of the measurements for a set of SME samples to account for the uncertainties of the measurements from this new analyzer.
【 预 览 】
附件列表
Files Size Format View
RO201704210001783LZ 88KB PDF download
  文献评价指标  
  下载次数:14次 浏览次数:16次