TA-55 Hot CMM Calibration Tolerance Analysis (U) | |
Montano, Joshua D.1  | |
[1] Los Alamos National Laboratory | |
关键词: ACCURACY; CALIBRATION; PERFORMANCE; SPECIFICATIONS; TOLERANCE; | |
DOI : 10.2172/1043515 RP-ID : LA-UR-12-22124 PID : OSTI ID: 1043515 Others : TRN: US201214%%141 |
|
学科分类:工程和技术(综合) | |
美国|英语 | |
来源: SciTech Connect | |
【 摘 要 】
The Hot Coordinate Measuring Machine (CMM), a Brown and Sharpe Xcel 765, has specifications listed by the manufacture of 4.5 + L/250 {micro}m for volumetric performance, 3.5 {micro}m for probing and 4.5 {micro}m for scanning. An upgrade was performed on the machine increasing its performance capability. This document reviews calibration data gathered after the upgrade over a five year period (2005-2010) and recommends a new specification of 3.0 + L/250 {micro}m for size, 3.3 {micro}m for probing, and 4.3 {micro}m for scanning. The new equations are an approximate 30% increase in accuracy for size and approximately 5% increase for probing and scanning.
【 预 览 】
Files | Size | Format | View |
---|---|---|---|
RO201704190002787LZ | 772KB | download |