科技报告详细信息
Final Report: A CdZnTe detector for MRI-compatible SPECT Systems
Meng, Ling-Jian
关键词: imaging;    MRI-compatible SPECT;    CdZnTe detector;   
DOI  :  10.2172/1058843
RP-ID  :  A CdZnTe detector for MRI-compatible SPECT Systems
PID  :  OSTI ID: 1058843
学科分类:化学(综合)
美国|英语
来源: SciTech Connect
PDF
【 摘 要 】

The key objective of this project is to develop the enabling technology for future MRI-compatible nuclear (e.g. SPECT) imaging system, and to demonstrate the feasibility of performing simultaneous MR and SPECT imaging studies of the same object. During the past three years, we have developed (a) a MRI-compatible ultrahigh resolution gamma ray detector and associated readout electronics, (b) a theoretical approach for modeling the effect of strong magnetic field on SPECT image quality, and (c) a maximum-likelihood (ML) based reconstruction routine with correction for the MR-induced distortion. With this support, we have also constructed a four-head MR-compatible SPECT system and tested the system inside a 3-T clinical MR-scanner located on UI campus. The experimental results obtained with this system have clearly demonstrated that sub-500um spatial resolution can be achieved with a SPECT system operated inside a 3-T MRI scanner. During the past three years, we have accomplished most of the major objectives outlined in the original proposal. These research efforts have laid out a solid foundation the development of future MR-compatible SPECT systems for both pre-clinical and clinical imaging applications.

【 预 览 】
附件列表
Files Size Format View
RO201704190000079LZ 21568KB PDF download
  文献评价指标  
  下载次数:18次 浏览次数:19次