SOLVENT HOLD TANK SAMPLE RESULTS FOR MCU-13-189, MCU-13-190, AND MCU-13-191: QUARTERLY SAMPLE FROM SEPTEMBER 2013
Fondeur, F. ; Taylor-Pashow, K.
关键词: MCU;    ARP;    ISDP;   
DOI  :  10.2172/1104731
RP-ID  :  SRNL-STI-2013-00652
PID  :  OSTI ID: 1104731
美国|英语
来源: SciTech Connect
PDF
【 摘 要 】

Savannah River National Laboratory (SRNL) analyzed solvent samples from Modular Caustic-Side Solvent Extraction Unit (MCU) in support of continuing operations. A quarterly analysis of the solvent is required to maintain solvent composition within specifications. Analytical results of the analyses of Solvent Hold Tank (SHT) samples MCU-13-189, MCU-13-190, and MCU-13-191 received on September 4, 2013 are reported. The results show that the solvent (remaining heel in the SHT tank) at MCU contains excess Isopar L and a deficit concentration of modifier and trioctylamine when compared to the standard MCU solvent. As with the previous solvent sample results, these analyses indicate that the solvent does not require Isopar L trimming at this time. Since MCU is switching to NGS, there is no need to add TOA nor modifier. SRNL also analyzed the SHT sample for {{sup 137}Cs content and determined the measured value is within tolerance and the value has returned to levels observed in 2011.