| Fabrication of Gamma Detectors Based on Magnetic Ag:Er Microcalorimeters | |
| Friedrich, Stephan1  Boyd, Stephen2  Cantor, Robin3  | |
| [1] Lawrence Livermore National Lab. (LLNL), Livermore, CA (United States);Univ. of New Mexico, Albuquerque, NM (United States);STAR Cryoelectronics, Santa Fe, NM (United States) | |
| 关键词: CALORIMETERS; DOPED MATERIALS; SQUID DEVICES; FABRICATION; GOLD; SILVER; ERBIUM; GAMMA DETECTION; SIGNAL-TO-NOISE RATIO; SENSORS; RESOLUTION; SENSITIVITY; ZEEMAN EFFECT; MAGNETIC FIELDS; TEMPERATURE RANGE 0000-0013 K; | |
| DOI : 10.2172/1259772 RP-ID : LLNL--TR-679717 PID : OSTI ID: 1259772 Others : TRN: US1601532 |
|
| 美国|英语 | |
| 来源: SciTech Connect | |
PDF
|
|
【 摘 要 】
This report discusses the photolithographic fabrication of ultra-high resolution gamma-ray detectors based on magnetic microcalorimeters (MMCs). The MMC uses a novel Er-doped silver sensor (Ag:Er) that is expected to have higher sensitivity than the Er-doped gold (Au:Er) sensors currently in use. The MMC also integrates the first-stage SQUID preamplifier on the same chip as the MMC gamma detector to increase its signal-to-noise ratio. In addition, the MMC uses a passive Ta-Nb heat switch to replace one of the common long-term failure points in earlier detectors. This report discusses the fabrication process we have developed to implement the proposed improvements.