科技报告详细信息
On Line Enrichment Monitor (OLEM) UF6 Tests for 1.5" Sch40 SS Pipe, Revision 1
March-Leuba, Jos?? A.1  Garner, Jim1  Younkin, Jim1  Simmons, Darrell W.1 
[1]Oak Ridge National Lab. (ORNL), Oak Ridge, TN (United States)
关键词: ON-LINE MEASUREMENT SYSTEMS;    HIGHLY ENRICHED URANIUM;    SLIGHTLY ENRICHED URANIUM;    DEPLETED URANIUM;    URANIUM HEXAFLUORIDE;    CENTRIFUGE;   
DOI  :  10.2172/1257899
RP-ID  :  ORNL/LTR--2015/773
PID  :  OSTI ID: 1257899
Others  :  TRN: US1601498
美国|英语
来源: SciTech Connect
PDF
【 摘 要 】
As global uranium enrichment capacity under international safeguards expands, the International Atomic Energy Agency (IAEA) is challenged to develop effective safeguards approaches at gaseous centrifuge enrichment plants while working within budgetary constraints. The ???Model Safeguards Approach for Gas Centrifuge Enrichment Plants??? (GCEPs) developed by the IAEA Division of Concepts and Planning in June 2006, defines the three primary Safeguards objectives to be the timely detection of: 1) diversion of significant quantities of natural (NU), depleted (DU) or low-enriched uranium (LEU) from declared plant flow, 2) facility misuse to produce undeclared LEU product from undeclared feed, and 3) facility misuse to produce enrichments higher than the declared maximum, in particular, highly enriched uranium (HEU). The ability to continuously and independently (i.e. with a minimum of information from the facility operator) monitor not only the uranium mass balance but also the 235U mass balance in the facility could help support all three verification objectives described above. Two key capabilities required to achieve an independent and accurate material balance are 1) continuous, unattended monitoring of in-process UF6 and 2) monitoring of cylinders entering and leaving the facility. The continuous monitoring of in-process UF6 would rely on a combination of load-cell monitoring of the cylinders at the feed and withdrawal stations, online monitoring of gas enrichment, and a high-accuracy net weight measurement of the cylinder contents. The Online Enrichment Monitor (OLEM) is the instrument that would continuously measure the time-dependent relative uranium enrichment, E(t), in weight percent 235U, of the gas filling or being withdrawn from the cylinders. The OLEM design concept combines gamma-ray spectrometry using a collimated NaI(Tl) detector with gas pressure and temperature data to calculate the enrichment of the UF6 gas within the unit header pipe as a function of time. The OLEM components have been tested on ORNL UF6 flow loop. Data were collected at five different enrichment levels (0.71%, 2.97%, 4.62%, 6.0%, and 93.7%) at several pressure conditions. The test data were collected in the standard OLEM N.4242 file format for each of the conditions with a 10-minute sampling period and then averaged over the span of constant pressures. Analysis of the collected data has provided enrichment constants that can be used for 1.5??? stainless steel schedule 40 pipe measurement sites. The enrichment constant is consistent among all the wide range of enrichment levels and pressures used.
【 预 览 】
附件列表
Files Size Format View
2738KB PDF download
  文献评价指标  
  下载次数:13次 浏览次数:28次