Fatigue resistance of nitinol stents implanted in femoropopliteal arteries is a critical issue because of their harsh biomechanical environment. Limb flexions due to daily walk expose the femoropopliteal arteries and, subsequently, the implanted stents to large cyclic deformations, which may lead to fatigue failure of the smart self-expandable stents. For the first time, this paper utilised the up-to-date measurements of walk-induced motion of a human femoropopliteal artery to investigate the fatigue behaviour of nitinol stent after implantation. The study was carried out by modelling the processes of angioplasty, stent crimping, self-expansion and deformation under diastolic-systolic blood pressure, repetitive bending, torsion and axial compression as well as their combination. The highest risk of fatigue failure of the niti -nol stent occurs under a combined loading condition, with the bending contributing the most, followed by compression and torsion. The pulsatile blood pressure alone hardly causes any fatigue failure of the stent. The work is significant for understanding and improving the fatigue performance of nitinol stents through innovative design and procedural optimisation. (C) 2021 The Author(s). Published by Elsevier Ltd.

【 授权许可】

Free   

【 预 览 】
附件列表
Files Size Format View
10_1016_j_jbiomech_2021_110295.pdf 3827KB PDF download
  文献评价指标  
  下载次数:16次 浏览次数:0次