期刊论文详细信息
JOURNAL OF BIOMECHANICS 卷:45
A hierarchy of computationally derived surgical and patient influences on metal on metal press-fit acetabular cup failure
Article
Clarke, S. G.1  Phillips, A. T. M.1  Bull, A. M. J.2  Cobb, J. P.3 
[1] Univ London Imperial Coll Sci Technol & Med, Dept Civil & Environm Engn, London SW7 2AZ, England
[2] Univ London Imperial Coll Sci Technol & Med, Dept Bioengn, Royal Sch Mines, London SW7 2AZ, England
[3] Univ London Imperial Coll Sci Technol & Med, MSk Lab, London W6 8RF, England
关键词: Orthopaedic surgery;    Cementless acetabular cup;    Modelling;    Hip replacement;    Finite element analysis;   
DOI  :  10.1016/j.jbiomech.2012.03.026
来源: Elsevier
PDF
【 摘 要 】

The impact of anatomical variation and surgical error on excessive wear and loosening of the acetabular component of large diameter metal-on-metal hip arthroplasties was measured using a multi-factorial analysis through 112 different simulations. Each surgical scenario was subject to eight different daily loading activities using finite element analysis. Excessive wear appears to be predominantly dependent on cup orientation, with inclination error having a higher influence than version error, according to the study findings. Acetabular cup loosening, as inferred from initial implant stability, appears to depend predominantly on factors concerning the area of cup-bone contact, specifically the level of cup seating achieved and the individual patient's anatomy. The extent of press fit obtained at time of surgery did not appear to influence either mechanism of failure in this study. (C) 2012 Elsevier Ltd. All rights reserved.

【 授权许可】

Free   

【 预 览 】
附件列表
Files Size Format View
10_1016_j_jbiomech_2012_03_026.pdf 461KB PDF download
  文献评价指标  
  下载次数:1次 浏览次数:0次