Use of beam probes for rigidity calibration of the A1900 fragment separator
Article; Proceedings Paper
Ginter, T. N.1  Farinon, F.2,3  Baumann, T.1  Hausmann, M.2  Kwan, E.1  Cuncic, O. Naviliat1  Portillo, M.2  Rogers, A. M.1,4  Stetson, J.1  Sumithrarachchi, C.1  Villari, A. C. C.2  Williams, S. J.1 
[1] Michigan State Univ, Natl Superconducting Cyclotron Lab, E Lansing, MI 48824 USA
[2] Michigan State Univ, Facil Rare Isotope Beams, E Lansing, MI 48824 USA
[3] GSI Darmstadt, Planckstr 1, D-64291 Darmstadt, Germany
[4] Univ Massachusetts Lowell, Dept Phys, Lowell, MA 01854 USA
关键词: Electromagnetic separator;    Rigidity;    Calibration;    Dipole magnet;    Fragmentation reactions;   
DOI  :  10.1016/j.nimb.2015.12.047
来源: Elsevier
PDF
【 摘 要 】

Use of a beam-based approach is presented for establishing a rigidity calibration for the A1900 fragment separator located at the National Superconducting Cyclotron Laboratory. Also presented is why an alternative approach to the rigidity calibration - using detailed field maps of individual magnetic components - is not a feasible basis for deriving an accurate calibration. The level of accuracy achieved for the rigidity calibration is +/- 0.1%. (C) 2016 Elsevier B.V. All rights reserved.