期刊论文详细信息
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION B-BEAM INTERACTIONS WITH MATERIALS AND ATOMS 卷:445
Evaluation of the continuous dilatometer method of silicon carbide thermometry for passive irradiation temperature determination
Article
Field, Kevin G.1  McDuffee, Joel L.1  Geringer, Josina W.1  Petrie, Christian M.1  Katoh, Yutai1 
[1] Oak Ridge Natl Lab, Oak Ridge, TN 37831 USA
关键词: Silicon carbide (SiC);    Passive irradiation temperature monitor;    Dilatometry;    Annealing;   
DOI  :  10.1016/j.nimb.2019.02.022
来源: Elsevier
PDF
【 摘 要 】

Silicon carbide (SiC) is a primary candidate for passive irradiation temperature monitoring, and continuous dilatometry (CD) has been proposed as the key method for extracting irradiation temperatures from SiC thermometry samples. The CD method was evaluated to determine the sensitivity of the technique to the analysis procedure, as well as its accuracy and precision. Analysis found that the CD method had a +/- 20 degrees C sensitivity to the algorithm used to determine the irradiation temperatures. Comparison of the CD method with the continuous length change method for SiC thermometry showed CD is a more viable method of extracting irradiation temperatures. A comparison of irradiation temperatures determined by CD with those from active thermocouple measurements further validated the CD method for accurately and precisely determining irradiation temperatures in materials test reactor irradiations.

【 授权许可】

Free   

【 预 览 】
附件列表
Files Size Format View
10_1016_j_nimb_2019_02_022.pdf 2147KB PDF download
  文献评价指标  